Degassing removes moisture, solvents and other contaminants from the surface and pores of the sample. If the treatment is insufficient, these contaminants block part of the surface, which results in an underestimated BET surface area. If the temperature is too high or the treatment too long, the material itself can change, for example through decomposition, dehydration, sintering or structural collapse. The degassing conditions must therefore be matched to the properties of the sample.